Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Determination of Water Vapor Permeability Coefficient: Principles, Standards, and Test Methods

Water vapor permeability coefficient determination is a core barrier-performance test that quantifies how much moisture migrates through plastic films, laminates, and functional materials under controlled temperature and humidity conditions. The results, together with water vapor transmission rate (WVTR) and permeance, directly govern shelf-life modeling, material selection, and compliance with standards such as ASTM E96/E96M, ASTM F1249, and ISO 15106. This article outlines the technical principles, specimen categories, mainstream test methods, and typical applications of water vapor permeability coefficient testing.

Technical Overview

The water vapor permeability coefficient is an intrinsic material property that describes the steady-state mass of water vapor passing through a material of unit thickness, across a unit area, per unit time, under a unit difference in water vapor partial pressure. It is commonly expressed in units of g·cm/(cm²·s·Pa) or ng·cm/(cm²·s·Pa). Because the coefficient is normalized by thickness, it allows direct comparison of the intrinsic moisture-barrier capability of different homogeneous materials, whereas transmission rate results are geometry dependent.

Three related quantities are routinely reported in moisture-barrier testing:

  • Water vapor transmission rate (WVTR): the mass of water vapor transmitted per unit area per unit time, typically in g/m²·24h, measured under specified temperature and relative humidity conditions.
  • Water vapor permeance: WVTR divided by the water vapor partial pressure difference across the specimen, expressed in g/(m²·s·Pa), used mainly for laminates and coated structures where a thickness-normalized coefficient is not physically meaningful.
  • Water vapor permeability coefficient: permeance multiplied by mean specimen thickness, expressed in g·cm/(cm²·s·Pa), applicable to single-layer homogeneous films and sheets.

These parameters are determined in accredited third-party laboratories using gravimetric (cup) methods or instrumented sensor methods, following recognized standards published by ISO, ASTM, and GB committees. Accurate determination requires strict control of temperature, humidity gradient, specimen sealing, and preconditioning, because water vapor transport in polymers is strongly dependent on both temperature and relative humidity.

Testing Samples

Plastic Films and Sheets

Single-layer and mono-material polymer films such as polyethylene (PE), polypropylene (PP), polyethylene terephthalate (PET), polyamide (PA), polyvinyl chloride (PVC), ethylene-vinyl alcohol copolymer (EVOH), polyvinylidene chloride (PVDC), and regenerated cellulose film. Specimens should be flat, wrinkle-free, and free of pinholes, with a test area and thickness representative of the production lot.

Flexible Packaging Laminates and Composites

Multi-layer structures produced by lamination, coextrusion, or metallization, including aluminum foil laminates, metallized PET and BOPP films, transparent high-barrier coated films such as SiOx and AlOx coated substrates, and printed flexible packaging. For these heterogeneous structures, laboratories usually report WVTR and permeance rather than a thickness-normalized coefficient.

Paper, Paperboard, and Textiles

Moisture-resistant papers, waxed or polymer-coated paperboard, release papers, nonwoven fabrics, and technical textiles used in packaging, medical dressings, and construction applications. These materials generally exhibit higher transmission rates and are often tested by the cup method under moderate humidity gradients.

High-Barrier and Functional Materials

Photovoltaic encapsulant films and backsheets, OLED and flexible-electronics barrier films, medical device packaging, sterile barrier systems, construction vapor-control membranes, breathable films, and geomembranes. These applications may demand transmission rates ranging from below 10⁻³ g/m²·24h for ultra-high-barrier encapsulation up to several hundred g/m²·24h for breathable structures, which dictates the selection of a measurement technique with adequate sensitivity.

Testing Items

Typical test items and reported results include:

  • Water vapor transmission rate (WVTR) in g/m²·24h at specified conditions, for example 38 °C / 90% RH or 23 °C / 85% RH.
  • Water vapor permeance in g/(m²·s·Pa).
  • Water vapor permeability coefficient in g·cm/(cm²·s·Pa) or ng·cm/(cm²·s·Pa).
  • Thickness measurement of the specimen at multiple points and calculation of the coefficient from the measured WVTR.
  • Comparative barrier ranking of candidate structures and verification of supplier datasheet values.
  • Temperature- and humidity-dependent WVTR curves for shelf-life and reliability modeling.

Each test report states the method, the standard applied, specimen orientation where relevant, conditioning, test conditions, number of specimens, and the mean result with standard deviation, since transmission behavior cannot be interpreted without its conditioning and test environment.

Testing Methods

Gravimetric (Cup) Method

Specimens are sealed over cups containing either a desiccant (desiccant method) or distilled water (water method) and placed in a controlled-humidity, controlled-temperature chamber. Water vapor diffusing through the specimen changes the cup mass, which is monitored gravimetrically until steady state is reached. The transmission rate is calculated from the slope of the mass-versus-time curve divided by the test area. Representative standards include ASTM E96/E96M, ISO 2528, GB/T 1037, and JIS Z 0208. The method is robust and covers a wide range of transmission rates, but it requires several days to reach equilibrium and manual weighing can limit repeatability at low transmission levels.

Infrared Sensor Method

The specimen separates a controlled wet chamber from a dry carrier-gas stream. Water vapor permeating the specimen is swept to an infrared detector whose signal is proportional to the water concentration in the gas. The instrument is calibrated with a reference film of certified transmission rate, after which WVTR is read directly once steady state is achieved. This technique underpins ASTM F1249, ISO 15106-2, and GB/T 26253, and is widely used for fast, high-throughput testing of high-barrier films with detection limits down to the 10⁻³ g/m²·24h level.

Electrolytic (Coulometric) Sensor Method

Permeated water vapor is carried by a dry carrier gas to an electrolytic cell, where it is quantitatively electrolyzed; the electrolysis current is a direct Faradaic measure of the water mass flow. Because the measurement is absolute, the technique offers excellent long-term accuracy and stability for medium-to-high barrier materials. It is standardized in ISO 15106-3, GB/T 21529, and related national standards, and supports fully automatic monitoring of steady state over extended test periods.

Humidity Sensor Method

A sealed test cell is divided by the specimen into a humidified side and a dry side equipped with a relative-humidity sensor. The rate of humidity rise on the dry side, together with the cell volume and temperature, yields the transmission rate. This principle is described in ISO 15106-1 and corresponding GB standards, and modern implementations provide rapid measurement well suited to quality control of medium-barrier materials.

Test Conditions, Preconditioning, and Calculation

Specimens are conditioned before testing, commonly at 23 °C and 50% RH. The most frequently used test conditions are 38 °C / 90% RH for general barrier films and photovoltaic materials, 23 °C / 85% RH or 23 °C / 50% RH for ambient comparisons, and 40 °C / 90% RH for certain pharmaceutical and medical applications; ultra-high-barrier encapsulation layers may be evaluated at 38 °C / 100% RH or under 85 °C / 85% RH damp-heat conditions. Once steady-state WVTR is obtained, the permeability coefficient is calculated by multiplying the measured permeance, that is WVTR divided by the water vapor partial pressure difference across the specimen at the test temperature, by the mean specimen thickness. Testing is normally performed on at least three specimens, and results are rejected if leakage is detected around the seal.

Applications

Determination of the water vapor permeability coefficient supports quality control, material development, and regulatory compliance across many industries:

  • Food and beverage packaging: matching film and laminate barrier classes to product moisture sensitivity for shelf-life assurance.
  • Pharmaceutical and medical packaging: qualifying blister films, sachets, sterile barrier systems, and desiccant-based designs.
  • Photovoltaics: verifying backsheets, encapsulant films, and edge-seal materials against module damp-heat reliability requirements, such as the 85 °C / 85% RH exposure protocols defined in IEC 61215.
  • Flexible and organic electronics: screening encapsulation films with ultra-low transmission rates for OLED and thin-film device protection.
  • Building and construction: selecting vapor-control layers, breathable membranes, and coatings with engineered moisture behavior.
  • Textiles and hygiene products: characterizing waterproof yet breathable membranes and laminates.
  • Research and development: comparing resin grades, coating chemistries, and processing parameters during barrier-material development.

FAQ

What is the difference between WVTR, permeance, and the permeability coefficient?

WVTR describes the transmission rate per unit area under stated conditions and depends on both the material and its thickness. Permeance is WVTR normalized by the vapor pressure difference across the specimen. The permeability coefficient additionally normalizes by thickness and therefore represents an intrinsic material property; it is meaningful only for homogeneous single-layer materials, whereas laminates and coated structures are characterized by WVTR and permeance.

Which method should be selected for a given sample?

The cup method is suitable for papers, textiles, and medium-barrier films, while infrared and electrolytic sensor methods are preferred for high-barrier films where speed and sensitivity matter. When a product standard or customer specification mandates a particular method and condition, that specification governs. Accredited third-party laboratories can advise on the most appropriate technique for the expected barrier range.

What are the typical test conditions and how long does a test take?

The most common condition is 38 °C with a 90% RH gradient. Cup-method tests typically require three to fourteen days to reach steady state, whereas sensor-based instruments often complete a test within hours for medium-barrier materials; ultra-high-barrier films may require extended equilibration to confirm the steady-state value.

How are specimens prepared?

Specimens must be flat, defect-free, and representative of the lot, sealed leak-tight into the test cell or cup, and conditioned before measurement. Thickness is measured at multiple points because it enters directly into the coefficient calculation. Sealing quality is monitored throughout the test, since edge leakage is the dominant source of error in moisture-barrier measurements.

Can results from different methods be compared directly?

Only with caution. Different standards specify different conditions, and gravimetric and sensor-based methods can exhibit method bias. Comparisons should always be made at identical temperature, humidity, and specimen orientation, and certified reference films are recommended for cross-verification between laboratories.

How are results reported?

Reports state the standard and method applied, conditioning, test temperature and humidity, specimen thickness, individual and mean WVTR values, and, where applicable, the calculated permeance and permeability coefficient with units, measurement uncertainty, and the number of specimens tested.